A vacuum shoebrush should be provided outside the dressing area for use for five
seconds on each. A tacky matt placed at the entrance of the dressing area will help remove
even more particles from shoes.
Opening Cleanroom Doors
Make sure all contamination sensitive parts and surfaces are moved away from the
doorway. Even though the flow of air should be to the outside, there is still a chance of
particles being injected into the cleanroom while doors are open.
Air Turbulence
In an empty cleanroom, a wall of filtered air travels unimpeded in either a vertical
or horizontal flow. Unfortunately, whenever anything is put in the path of the air flow,
including equipment or people, a turbulent air pattern called a vortex is created. Much
like dust in a whirlwind, particles tend to be drawn into these turbulent areas, and
resettle throughout the cleanroom. It is imperative that equipment and people be stationed
as far away from the direct face or flow of air in the cleanroom as possible.
Air Pressure
The air pressure in the cleanroom should be positive in relation to adjacent rooms.
This will help to prevent the inflow of particles from the outside when doors are opened.
Air pressure can be adjusted through the cleanroom air handling system.
Static Control
When the relative humidity drops below 30% in the cleanroom, there is very little
moisture in the air to dissipate static charges. These static charges are capable of doing
physical damage to sensitive electronic parts. As these charges accumulate on furniture,
equipment, people and critical products, particles will also be attracted to these
surfaces. Static can be controlled through the cleanroom air handling system by adjusting
the relative humidity of the incoming air. People working in the cleanroom can also wear
grounded wrist straps to help dissipate static charges.
Product Handling
Never use bare hands to touch a critical product. Hands contain body oils as well as
dirt and skin particles that can be transferred to the product surface. Always wear
approved gloves or use clean handling tools such as tweezers or tongs.
Maintenance/Cleaning
The cleanroom must be cleaned on a regular schedule determined by the contamination
control engineer, who should keep a log book of each cleaning. With few exceptions, it
should be cleaned either before or after the regular working shift, with the floors and
all horizontal surfaces being the main priority. Walls and ceilings may require cleaning
after extended shutdowns or degradation of air cleanliness.
Tape the wheels on all equipment and furniture with cleanroom approved
tape. If an item entering the cleanroom cannot be cleaned, it should be sealed in approved
packaging material.
Wipe down all incoming equipment with cleanroom approved wipers and
clean-filtered solvent using a unidirectional wiping stroke rather than a back and forth
motion.
Water that has been deionized and filtered can serve as an effective
cleaning solution in the cleanroom. Do not used powdered detergents but only approved
liquid detergents. Rinse thoroughly to remove all detergent residue.
Only use wipers with low particle and low residue properties. Avoid
tissue type wipers. Also avoid wipers that have been treated with surfactants, as they can
leave behind residues and react with process solvents to produce even more contaminants.
Hardware Packaging
All contamination sensitive parts leaving the clean room should be packaged in
approved cleanroom packaging material and sealed using a heat sealer or cleanroom approved
tape. Do not put two items in the same bag, as they will generate particles when they
touch or rub together. Open sealed packages with a razor blade or by sliding scissor
blades clear through.
Storage
Contamination sensitive parts should be stored in protective packaging or an
ultraclean container and placed out of the flow of cleanroom traffic.
Shutdown
When the cleanroom is shut down due to a power outage or major maintenance, the room
must be recertified to meet Federal Standard 209D Cleanliness Class requirements. A down
time of less than 72 hours requires only routine cleaning. For periods of over 72 hours,
the cleanroom should be cleaned from ceiling to floor before recertification.
Safety
Know where all fire exits and fire extinguishers are located in the cleanroom. Be
aware of any hazardous chemicals or substances in your vicinity and know how to deal with
any accidents that may occur.